Operation Manual
6 GMC-I Messtechnik GmbH
3 Initial Start-Up
3.1 Battery Test
Five battery symbol segments ranging from depleted to fully charged
continuously indicate the current battery level at the main menu.
3.2 Installing and Replacing Batteries
New batteries must be installed before initial start-up, or when only one
solid segment remains in the battery symbol.
Attention!
!
The instrument must be disconnected from the measuring circuit
(mains) at all poles before the battery compartment is opened.
Four 1.5 V baby cells in accordance with IEC LR14 are required for
operation of the METRISO C. Use alkaline-manganese batteries only.
Rechargeable NiCd or NiMH batteries can be used as well. Be absolutely
sure to refer to chapter 13.2, page 26, regarding the charging cycle and
the charging adapter.
Always replace the batteries in complete sets.
Dispose of batteries in an environmentally sound fashion.
Ð Loosen the two slotted screws at the battery compartment cover on
the housing rear panel and remove the cover.
Ð Insert four 1.5 V baby cells making certain they are poled in
accordance with the symbols. Insert the two batteries which are half
covered by the housing first.
Ð Replace the cover and retighten the screws.
Attention!
!
The instrument may not be operated if the battery compartment
cover has not been installed and properly tightened!
3.3 User Guide in a Different Language
By performing a software update it is possible to load another language
for the user guide than the one included in the scope of supplies. Any lan-
guage currently available is proposed for choice when WinProfi is being
installed, see chapter 3.5.
3.4 Selecting a Menu and Configuring Basic Settings
Press the or the key in order to display the desired measuring
function, the device settings or the database functions.
START
START