Operation Manual
10 GMC-I Messtechnik GmbH
B Prerequisites for Software Update or Data Exchange
Ð Find the interface to which the METRISO C is connected.
Note
+
Always start this function first, before performing an update or chang-
ing report templates. After starting this function, WinProfi loads the
report files specifically necessary for the connected instrument.
Due to the fact that WinProfi has been created for use with several
types of test instruments, incorrect test reports may otherwise be
loaded, or erroneous options may be made available.
Ð Query information regarding current software version
C Transmission of a Software Update to the Test Instrument
Ð PC: Select the Update All function from the Update menu. Follow the
instructions which appear at the monitor.
Depending upon the utilized PC, transmission takes from 1 to 2 minutes.
The NETZ/MAINS LED of the METRISO C test instrument lights up green and indicates
that the instrument is ready to receive data. If the PC and the test instrument are correctly
synchronized, the same LED lights up yellow. During programming sequences, the LIMIT
and U>25V LEDs light up red and the NETZ/MAINS LED lights up yellow in alternating
order. Upon completion of data transmission, the NETZ/MAINS LED briefly lights up green,
afterwards all LEDs go out.
The message „Transmission done“ appears on the computer screen
Attention!
!
The instrument may not, under any circumstances, be switched
off during transmission, nor may the connection between the in-
strument and the PC be interrupted!