3.4 Checking Device Functions
The METRISO 5024 functions as a voltmeter when the toggle switch 
(4) is in the neutral position:
Ð The voltmeter function can be checked by applying a voltage to 
both test probes (1 and 8), e.g. line voltage from a mains outlet.
The rotary switch must be set to any of positions 100 V, 250 V or 
500 V.
Ð When the toggle switch (4) is in either the Ω or MΩ position, the 
batteries are activated and the METRISO 5024 remains in the 
selected function as long as the toggle switch is activated. 
Measurements in the Ω and MΩ ranges are evaluated by means 
of the green LED (10), the beeper and the pointer.
Ð The ohmmeter function – rotary switch in the Ω position and toggle 
switch in the Ω position – can be tested by short-circuiting the 
two test probes (1 and 8): 0 Ω appears at the display, 
continuous acoustic signal from the beeper.
Avoid open test probes (resistance approaching ∞), (pointer 
oscillates, periodic sequence of acoustic signals from the 
beeper).
Ð The insulation measuring function – rotary switch in the xxx V 
position and toggle switch in the MΩ position – can be tested 
with the help of the two limit values: Short-circuit the two test 
probes (1 and 8): Pointer oscillates, periodic sequence of 
acoustic signals from the beeper, green LED blinks; separate the 
test probes: ∞ appears at the display, continuous acoustic signal 
from the beeper, green LED lights up.
Display accuracy can be tested using the ISO Calibrator 1 (article no. 
M662A) as an accessory by connecting it to the low or high value 
resistor.