Operation Manual
Table Of Contents

GMC-I Messtechnik GmbH 5
uppermost scale, low-resistance at the middle scale and voltage at 
the bottom scale. The device’s operating status and battery charge 
level are indicated with the 2-color LED in the scale (Figure 1).
The battery compartment is located in the housing base (6). The 
procedure for replacing batteries is described in 3.2.
All measurements are performed with two test probes:
• One test probe (1) is permanently attached to the housing – 
positive pole.
• The other (8) is attached to a coil cord – negative pole
This test probe (8) can be inserted into the opening (9) and secured 
by snapping the handle into the detaining fork (7).
As long as the toggle switch (4) is in the neutral middle position, the 
METRISO 5024 can be utilized as a voltmeter. 
The rotary switch must be set either to the 100 V, 250 V or 500 V 
position in order to perform voltage measurements. No batteries are 
required for this function.
The rotary switch must be set to the Ω position in order to perform 
low-resistance measurements. Measurement is performed with the 
toggle switch (4) in the Ω position. Voltage cannot be measured as 
long as the rotary switch (3) remains in this position.
The desired test voltage must be selected with the rotary switch (3) 
before performing insulation measurements. Measurement is 
performed with the toggle switch (4) in the MΩ position.
3 Preparation for Device Operation
3.1 Unpacking the Device
Check for mechanical damage after unpacking the device. The 
device is ready for use as soon as the batteries have been inserted 
into the battery compartment.
3.2 Power Supply – Battery Replacement
The METRISO 5024 is operated exclusively with batteries for safety 
reasons. Batteries are inserted or replaced as follows:
Ð First disconnect the measurement cables from the device under 
test.
Ð Loosen the fastening screw at the bottom of the battery 
compartment (6) and remove the battery compartment.
Ð Remove the depleted batteries and replace them with new or 
fully charged batteries. Make sure that battery polarity is not 
reversed during insertion (see printed symbols on the 
compartment).
Ð Push the battery compartment back in and secure it with the 
screw.
Recommended batteries: 4 ea. alkaline batteries per LR6, mignon 
cells, AA cells, 1.5 V, at least 2300 mAh
3.3 Checking the Mechanical Zero Point
Ð Lay the device horizontally onto a work surface.
Ð Check the pointer for correct zero point adjustment and reset 
with the adjusting screw (2a) if necessary. 
The device may not be connected to a measuring circuit during 
this procedure, and the toggle switch may not be activated. 
Don’t forget: Negative pointer deflection is suppressed by the 
scale’s limit stop. Always turn the adjusting screw clockwise for 
this reason, and then slowly turn it counterclockwise until the 0 Ω 
or 0 V position is reached.