User Manual
GOSSEN METRAWATT GMBH 29
Electrical Safety
Protection Class II
Nom. Mains Voltage 230 V
Test Voltage Mains + PE (mains) + 2 mA socket for testing for the absence of
voltage at test socket, connector jacks for phase and protective
conductors and gripper clip:
3kV∼ mains to PE (mains) + 2 mA socket: 1.5 kV∼
Measurement Category II
Contamination Level 2
Safety Cut-Off when device overheats
Electromagnetic Compatibility EMV
Product standards EN 61326-1: 1997
EN 61326: 1997/A1: 1998
Ambient Conditions
Operation − 10 ... + 55 °C
Storage − 25 ... + 70 °C
Atmosph. Humidity max. 75%
Elevation to 2000 m
Mechanical Design
Protection Device IP 40, terminals IP 20
Dimensions W x H x D: 190 mm x 140 mm x 95 mm
Weight 1.3 kg
12 Maintenance
No special maintenance is required. Keep outside surfaces clean and dry. Use a slightly
dampened cloth for cleaning. Avoid the use of solvents, cleansers or abrasives.
Interference Emission Class
EN 55022 A
Interference Immunity Test value Performance Feature
EN 61000-4-2 Contact/atmos. – 4 kV/8 kV B
EN 61000-4-3 10 V/m B
EN 61000-4-4 Mains connection – 2 kV B
EN 61000-4-5 Mains connection – 1 kV A
EN 61000-4-6 Mains connection – 3 V B
EN 61000-4-11 0.5 period / 100% A