8 Measuring Load Current from the Mains Outlet
Attention!
!
The device under test may not be connected to the mains outlet (7) until is has
passed safety testing in accordance with DIN VDE 0701, part 1!
➭ Connect the earthing contact plug from the device under test to the mains outlet (7).
➭ Set the function selctor switch to the “I
Netz
16 A” position.
➭ Switch the device under test on.
➭ Read the measured value in “A” from the LCD (11).
Attention!
!
Maximum allowable load capacity is 16 A continuous and 20 A for up to 10 minutes.
The electrical system to which the test instrument is connected must be protected
against overload with a fuse or circuit breaker. The fuse or circuit breaker rating may
not exceed 16 A!
To the mains outlet
I
Netz