Datasheet
28 GMC-I Messtechnik GmbH
8.2 Technical Data for Measurement Cables
(scope of delivery: KS17-2 safety cable set)
Electrical Safety
Ambient Conditions (EN 61010-031)
Temperature –20 C ... + 50 C
Relative humidity 50 to 80%
Contamination level
2
Application KS17-2
Attention!
!
Please observe the maximum values of the electrical safety of the
device.
In conformity with standard DIN EN 61010-031, measure-
ments in an environment according to measuring category III
may only be performed with the safety cap applied to the test
probe of the measurement cable.
For establishing contact in 4 mm jacks you have to remove the safety
cap by levering out the snap lock of the safety cap with another sharp
object (e.g. the second test probe).
Maximum Rated Voltage
600 V 1000 V
1000 V
Measuring Category
CAT IV CAT III CAT I I
Maximum Rated Current 1 A 1 A 16 A
with safety cap applied —
without safety cap applied — —