Datasheet
GMC-I Messtechnik GmbH 23
Fuse
Range up to 1 A and
FF 1.6A/600V, 6 mm x 32 mm protects the
10 μA ... 1 A and
x1 ... k x10 ranges in
conjunction with the power diodes
Switching capacity:
50 kA with 600 V~ (one half wave on ohmic load) overload limit of the
melting fuse in conjunction with the power diodes: max. 200 A with
600 V (one half wave on ohmic load)
10 A range FF 16A/600V AC, 6 mm x 32 mm,
Switching capacity 50 kA at 600 V~
Elektrical Safety
Protection class II per IEC/EN 61010-1:2001/VDE 0411-
1:2002
Measuring category II III
Operating voltage 600 V 300 V
Pollution degree
2
Test voltage 3.5 kV~ per IEC/EN 61010-1:2001/
VDE 0411-1:2002
Electromagnetic Compatibility (EMC)
Interference emission
EN 61326-1:2006 Klasse B
Interference immunity
EN 61326-1:2006
EN 61326-2-1:2006
Mechanical Design
Protection type Housing: IP 40, connection sockets: IP 20
Dimensions 146 x 118 x 44 mm (in closed condition)
Weight approx. 0.45 kg without battery