Influencing variables and rated operating ranges
Temperature +5 ... +23 ... +35 C
Limit temperatures for accuracy +5 ... +35 C
for operation 0 ... +40 C
for storage –25 ... +65 C
(without battery)
Position additional influence effect max. 1% of scale
length when the scale is inclined to the horizontal
between 0 and
120
Frequency additional influence effect
max.
10% of full scale value in the ranges
Auxiliary voltage No additional influence effect; the upper range
value for each measuring range must be set via
the potentiometer in the case of resistance mea-
surements.
Series-mode interference
voltage attenuation on V
> 60 dB with AC 50 Hz
on V
~ > 120 dB with DC
Common-mode
voltage attenuation > 120 dB with DC and 50 Hz AC
Composite voltages and
composite currents see chapter 4.5
Other influencing
variables per IEC 60051/EN 60051
Power Supply
Battery nominal voltage 9 V ;
9 V flat cell battery per IEC 6LR61 (6F22),
alkaline-manganese or NiCd/NiMH-Akku.
Service life with alkaline-manganese cell: approx. 1000 hours
with NiCd storage battery: approx. 200 hours
in the
x 1 range: ¼ of the entire service life
Battery test indication must be within battery test zone
100 mV; 3 V ... 600 V: 15 Hz ... 1 kHz
10 μA ... 10 A: 15 Hz ... 1 kHz
300 mV; 1 V: 15 Hz ... 200 Hz