Datasheet

20 GMC-I Messtechnik GmbH
5 Technical Data
1)
T he voltage measuring ranges 100 mV .... 10 V can also be used for current measurements
as shown in the following table (e.g. reverse currents or leak currents). The accuracy of these
measuring ranges then corresponds to class 5.
Frequency range on ~: 15 ... 100 Hz
Overload protection same as on the voltage measuring ranges.
2)
Measuring value = indicated value + constant ; 0 dB 0.775 V, i.e. 1 mW at 600
3)
protected by PTC thermistor; cooling time of thermistor before the start of next measure-
ment: grant at least 2 minutes.
Voltage
1)
Level
2)
Input
Resistance R
i
Overload
Protection
up to
3)
Const. Measuring Span
100 mV 20 dB 40 ... –18 dB
10 M / /50 pF
600 V
300 mV 10 dB 30 ... 8 dB 600 V
1 V 0 dB 20 ... + 2 dB 600 V
3 V +10 dB 10 ... + 12 dB 600 V
10 V +20 dB 0 ... + 22 dB 600 V
30 V +30 dB +10 ... + 32 dB 600 V
100 V +40 dB +20 ... + 42 dB 600 V
300 V +50 dB +30 ... + 52 dB 600 V
600 V +60 dB +40 ... + 62 dB 600 V
Range Current at Full Scale Value R
i
100 mV 10 nA
10 M
300 mV 30 nA
1 V 100 nA
3 V 300 nA
10 V 1 μA
30 V 3 μA
Current Input Resistance R
i
Voltage Drop U
Overload Protection up to
10 μA 10,0 k 100 mV 600 V
3)
100 μA 1,0 k 100 mV 600 V
3)
1 mA 100,0 100 mV 600 V
3)
10 mA 10,0 100 mV 600 V
3)
100 mA 1,4 140 mV 600 V
3)
1 A 480 m 0,480 V 600 V
3)
1 0 A 26 m 260 mV 600 V
4)