Datasheet

GMC-I Messtechnik GmbH 15
Ð Select the measuring range of the multimeter as a function of the
measuring range and the transformation ratio of the clip-on trans-
former.
Ð Connect the measurement cables as shown.
Ð Set switch to position „ON“.
Ð Scan the electrical circuit with the clip-on transformer.
Ð Read the measured value by making allowances for the transforma-
tion ratio of the transformer.
Ð Disconnect the clip-on transformer from the electrical circuit before
separating the measurement cables from the instrument!
4.4.5 DC and AC Currents with (Clip-on) Current Sensors
(Current Transformers with Voltage Output)
Ð Set the range switch to position V or V~.
Ð Select the measuring range of the multimeter as a function of the
measuring range and the transformation ratio of the clip-on sensor.
Ð Connect the measurement cables as shown.
Ð Set switch to position „ON“.
Ð Scan the electrical circuit with the clip-on sensor.
Ð Read the measured value by making allowances for the transforma-
tion ratio of the transformer.
Example Current Sensor Z13B:
Measuring range of clip-on sensor: 60 A /40 A~,
Output of clip-on sensor: 10 mV/A (max. 600 mV)
Measuring range of multimeter: 1 V /~ 300 mV /~
Indicated value: 150 mV 15.0 A
V/~
I(A)
Load