Datasheet
10 GMC-I Messtechnik GmbH
Checking the battery voltage
If the pointer is within the zone marked with the battery symbol „ “
after the instrument has been turned on then the battery voltage is suf-
ficient that is, it is assured that the error limits according to the data
given in chapter 5 „Technical Data“ are maintained.
4.3 Voltage Measurement
Attention!
!
Regardless of the magnitude of the voltage being measured,
for safety reasons, the sum of the voltage being measured
and the voltage to earth must not exceed 600 V CAT II or
300 V CAT III when the multimeter is connected directly!
If possible, take the connection socket, marked „“, immediately to
earth or to the point having the lowest potential to earth when making
any voltage measurements.
4.3.1 DC and AC Voltages up to 600 V (Direct Connection)
Ð Set the range switch to posi-
tion 600 V or 600 V~.
Ð Connect the measurement
cables as shown.
With DC voltage note the po-
larity! The „“ jack must be
grounded. For safety rea-
sons, leads with connection
plugs that are protected against accidental contact (KS17-2) should
be used.
Ð Set switch to position „ON“.
Ð Work down to lower voltage ranges until optimum indication is ob-
tained.
Ð Read the measuring voltage on the VA scale.
Note: Overload protection up to 600 V is provided by a PTC thermis-
tor; cooling time of the thermistor until the start of next measurement:
grant at least 2 minutes.
– / ~ + / ~