4 Technical Data
4.1 Individual Device Functions
Fault Loop Impedance / Line Impedance
Nominal Range per EN 61557-3: 0.27 to 200
Voltage range: 190 to 260 V / 48 to 52 Hz
Load resistance:50 (variable number of pulses at 10 ms)
Fault Loop Impedance Without Tripping the RCCB
Nominal Range per EN 61557-3: 0.8 to 200
Voltage range: 190 to 260 V / 48 to 52 Hz
Load resistance:50 (variable number of pulses and pulse width)
Short-Circuit Current
Alternating Voltage (TRMS)
Frequency range: 48 to 52 Hz
1
Display for brief representation
2
Display for detailed representation
Key:
a) The measuring uncertainties specified here for fault loop impedance, line imped-
ance and short-circuit current are only valid if line voltage is stable during measure-
ment and if no other electrical circuits parallel to the measured circuit are in use.
c) rdg. means reading, i.e. measured value, D = digits (i.e. number of the decimal
place with the least significance)
4.2 General Data
Reference Conditions
Temperature 23 ± 2° C
Relative humidity 40 to 60%
Line voltage 230 V
2% / 50 Hz 1%
Device position any
Ambient Conditions
Operating Conditions
Operating temperature
0 to 40° C
Relative Humidity max. 85 %, no condensation allowed
Line voltage 190 to 260 V / 48 to 52 Hz
Device position any
Storage Conditions
Temperature -10 to 70° C
Relative Humidity max. 90% at -10 to +40° C
max. 80% at +40 to +70° C
Device position any
Power Supply
Batteries 4 ea. AAA (LR03), 1.5 V alkaline or
1.2 V NIMH (with at least 750 mAh)
Number of
measurements with batteries at 800 mAh:
approx. 3,000 measurements
Electrical Safety
Measuring category CAT III / 300 V
Pollution degree 2
Protection class II
Fuse SIBA ceramic fuse
6.3 mm x 32 mm, F1 A/600 V
switching capacity 50 kA at 600 V
Mechanical Design
Display OLED, multicolored, graphic
Protection IP 43
Dimensions approx. 260 x 70 x 40 mm
Weight approx. 0.36 kg with batteries
Range Resolution Intrinsic Uncertainty Measuring
Uncertainty
0.00 to 4.99 0.01 (3% rdg. + 5 D) (4% rdg. + 7 D)
5.0 to 49.9 0.1 (3% rdg. + 3 D) (4% rdg. + 4 D)
50 to 200 1 3% rdg. 4% rdg.
Range Resolution Intrinsic Uncertainty Measuring
Uncertainty
0.0 to 4.9 0.1 (5% rdg. + 2 D) (6% rdg. + 2 D)
50 to 200 1 7% rdg. 8% rdg.
Range Resolution Intrinsic Uncertainty Measuring
Uncertainty
0 to 999 A 1 A
Depending on
measuring error
for loop impedance
1 D
Depending on
measuring error
for loop impedance
1 D
1.0 to 9.9 kA 0.1 kA
10 to 23 kA 1 kA
Range Resolution Intrinsic Uncertainty Measuring
Uncertainty
24 to 260 V 1 V
1
0.1 V
2
(2% rdg. + 2 D) (3% rdg. + 3 D)