User Manual

GMC-I Messtechnik GmbH 5
3.3 Measurement of Fault Loop Impedance and Line Imped-
ance
3.3.1 Measurement in Circuits without RCD – ~ Function
The “~” function is suitable for measuring fault loop impedance in
electrical circuits without RCCBs, as well as line impedance.
After switching the instrument on, connect it between L and
PE in order to measure fault loop impedance, and between L
and N to measure line impedance.
After the voltage value has settled in, briefly press the START
key in order to trigger measurement. Assure good contact be-
tween the test probes and the device under test while mea-
surement is being performed!
Results after completion of the measurement:
Key:
Z Measured impedance ()
Z x 1.5 Multiplier for measured impedance Z ()
I
K Short-circuit current calculated as: I
SC
= 230 / (Z x 1.5) (A)
Remove the instrument.
3.3.2 Displaying Further Measured/Calculated Values
Stated briefly: The following are displayed, one after the other, by
pressing the DISP
and DISP keys:
Short-circuit current
1.5 times measured impedance
Impedance corrected by a value equal to measuring error
Measured impedance
In detail: Impedance corrected by a value equal to measuring error
is displayed by pressing the DISP
key, and only the measured
value without correction is displayed by pressing the DISP
key.
3.3.3 Measurement in Electrical Circuits with RCD
Testing Without Tripping the RCCB
Select the “RCD” function if loop impedance needs to be mea-
sured via an RCCB, without causing it to trip.
After switching the instrument on, connect it between L and N
in order to measure line impedance.
After the voltage value has settled in, briefly press the ~/RCD
key, which starts measurement of line impedance without trip-
ping the RCD. This is important if loop impedance needs to be
measured via an RCCB.
Assure good contact between the test probes and the device
under test while measurement is being performed!
The results appear at the display, and the “~” symbol is replaced
with the “RCD” symbol at the same time.
Figure 3.12: Connection Example: Loop Impedance Measurement, L2–PE
Figure 3.13: Sample Results for Imped-
ance Measurement (brief
representation)
Figure 3.14: Sample Results for Imped-
ance Measurement
(detailed representation)
Figure 3.15: Short-Circuit Current Figure 3.16: 1.5 Times Impedance
Figure 3.17: Impedance + Measuring
Error
Figure 3.18: Measured Impedance
Figure 3.19: Measured Impedance +
Measuring Error
Figure 3.20: Measured Impedance