Datasheet

3-349-690-03
2/10.12
GMC-I Messtechnik GmbH
METRALINE ISO
CHECK
Insulation Measuring Instrument
Ramp Function
Measurement
Table Display
Results
Applications
Measurement of insulation resistance at voltage-free devices
and systems, up to 1000 V depending upon variant
Measurement of surge protection devices, up to 1000 V
depending upon variant
Checking of test objects for absence of voltage
Applicable Regulations and Standards
Characteristic Values
Measurement of Insulation Resistance
Nominal Range per EN 61557-2: 0.100 M – Rmax*
* The Rmax value depends on the selected test voltage:
Nominal voltage of 50 to 99 V Rmax = 1.999 G
Nominal voltage of 100 to 249 V Rmax = 3.999 G
Nominal voltage of 250 to 1000 V Rmax = 9.999 G
Nominal measuring
voltage 50 to 1000 V
adjustable in steps of 1 V
Measuring voltage –0%/+10% of nominal voltage
Nominal measuring
current 1 mA (where Umes > Unom)
Short-circuit current < 3 mA
Automatic discharging
of the DUT Yes
Number of
measurements approx. 250
(with new alkaline batteries)
IEC 61010-1/-031
DIN EN 61010-1/-031
VDE 0411-1/-031
Safety requirements for electrical equipment for mea-
surement, control and laboratory use
Part 1: General requirements
Part 31: Safety requirements for hand-held probe as-
semblies for electrical measurement and test
IEC 61557-1/-2
DIN EN 61557-1/-2
VDE 0413-1/-2
Electrical safety in low voltage distribution systems up to
1000 V AC and 1500 V DC – Equipment for testing,
measuring or monitoring of protective measures
Part 1: General requirements
Part 2: Insulation resistance measuring
instruments
IEC 61326-1
DIN EN 61326-1
Electrical equipment for measurement, control and
laboratory use – EMC requirements
Part 1: General requirements
DIN EN 60529
VDE 0470-1
Degrees of protection provided by enclosures (IP code)
Range Reso-
lution
Intrinsic
Uncertainty
Measuring
Uncertainty
0.100 to 9.999 M 0.001 M (2% rdg. + 10 d) (3 % rdg. + 20 d)
10.00 to 99.99 M 0.01 M (2% rdg. + 10 d) (3 % rdg. + 20 d)
100.0 to 999.9 M 0.1 M (2% rdg. + 10 d) (3 % rdg. + 20 d)
1.000 G ... Rmax* 0.001 G (4 % rdg. + 15 d) (5 % rdg. + 25 d)
Insulation resistance measurement
with test voltages of 50 to 1000 V
Voltage measurement up to 600 V
Measurement of surge protection devices
with test voltages of 50 to 1000 V
Table of common varistors can be displayed
Digital display, backlit color OLED display
Indication of dangerous touch voltage
LED for measurement point illumination
Patented means of securing test probes