Safety Precautions
• The multimeter may only be operated by persons who are ca-
pable of recognizing touch hazards and taking the appropriate
safety precautions. Touch hazards according to the standard
exist anywhere, where voltages of greater than 33 V TRMS or
70 V DC may occur. Avoid working alone when taking mea-
surements which involve touch hazards. Be certain that a sec-
ond person is present.
• Maximum allowable voltage between the voltage measuring
sockets or all connector sockets and ground is 600 V for mea-
suring category III and 300 V for measuring category IV.
• Be aware of the fact that dangerous voltage peaks with signifi-
cant frequency components of greater than 1 kHz are not dis-
played when the low-pass filter is activated. We recommend
measuring voltage without the low-pass filter first, in order to
be able to detect any dangerous voltages.
• Be prepared for the occurrence of unexpected voltages at de-
vices under test (e.g. defective devices). For example, capaci-
tors may be dangerously charged.
• Make certain that the measurement cables are in flawless con-
dition, e.g. no damage to insulation, no interruptions in cables
or plugs etc.
• No measurements may be made with this instrument in electri-
cal circuits with corona discharge (high-voltage).
• Special care is required when measurements are made in HF
electrical circuits. Dangerous pulsating voltages may be pres-
ent.
• Measurements under moist ambient conditions are not permit-
ted.
• Be absolutely certain that the measuring ranges are not over-
loaded beyond their allowable capacities. Limit values are in-
cluded in section 8, “Technical Data”, in the table entitled
“Measuring Functions and Measuring Ranges” in the “Over-
load Capacity” column.
• The multimeter may only be operated with installed batteries or
rechargeable batteries. Dangerous currents and voltages are
otherwise not indicated, and the instrument may be damaged.
• Observe visual and acoustic warning signals!
• The instrument may not be operated if the fuse cover or the
battery compartment lid has been removed, or if its housing is
open.
• The input for the current measuring range is equipped with a
fuse link.
Use specified fuses only (see page 63)! The fuse must have a
breaking capacity of at least 30 kA.