Maintenance – Calibration
9.6 Recalibration
The measuring tasks performed with your instrument, and the
stressing it’s subjected to, influence aging of its components and
may result in deviation from the specified levels of accuracy.
In the case of strict measuring accuracy requirements, as well as
in the event of use at construction sites with frequent stress due
to transport and considerable temperature fluctuation, we recom-
mend a relatively short calibration interval of once per year. If your
instrument is used primarily in the laboratory and indoors without
considerable climatic or mechanical stressing, a calibration inter-
val of once every 2 to 3 years is sufficient as a rule.
During recalibration at an accredited calibration laboratory
(DIN EN ISO/IEC 17025), deviations from traceable standards
demonstrated by your measuring instrument are documented.
Ascertained deviations are used to correct displayed values dur-
ing later use of the instrument.
We would be happy to perform DAkkS or factory calibration for
you at our calibration laboratory. Further information is available at
our website:
www.gossenmetrawatt.com (
Company
DAkkS Calibration
Center or
FAQs
Calibration Questions and Answers)
.
Recalibration of your instrument at regular intervals is essential for
the fulfillment of requirements according to quality management
systems per DIN EN ISO 9001.
* Examination of the specification, as well as adjustment, are not included
in calibration. However, in the case of our own products, any required
adjustment is performed and adherence to the specification is con-
firmed.
9.7 Manufacturer’s Guarantee
All METRA HIT measuring and calibration instruments are guaran-
teed for a period of 3 years after date of shipment. The manufac-
turer’s guarantee covers materials and workmanship. Damages
resulting from use for any other than the intended purpose, as
well as any and all consequential damages, are excluded.
Calibration is guaranteed for a period of 12 months.