Device and Measuring Parameters
Storage of Individual Values using the SAMPLE or dAtA Sampling Rate
If only manually selected values need to be saved, SAMPLE must
be selected as the StorE > rAtE sampling value. If memory mode
operation is then started, a single measured value is saved to per-
manent memory with time stamp when the DATA/MIN/MAX key is
pressed and held until two rapidly repeating acoustic signals are
generated.
If dAtA is selected as the StorE > rAtE sampling rate, and if mem-
ory mode operation is then started, measured values ascertained
with activated DATA function are automatically saved to perma-
nent memory with time stamp.
triG – Trigger Conditions (parameters for memory mode operation)
The StorE > triG SEt = Sto-ou / Sto-in / OFF setting can be used
to specify how measured value recording is started and stopped:
• triG = off: Recording is started with Store > Start and ended
with Store > Stop.
• triG = sto-ou: Recording is started as soon as a measured value
occurs which is outside of the selected measuring limits, and is
stopped as soon as the measuring limits are once again com-
plied with, or the selected recording period has elapsed.
• triG = sto-in: Recording is started as soon as a measured value
occurs which is within a specified band, and is stopped as
soon as this is no longer the case, or after the maximum re-
cording period has elapsed.
The band is specified with the help of the L_triG lower limit and
the H_triG upper limit. Querying takes place in the event that
triG OFF is selected. Band widths are specified in digits, and
30,000 digits corresponds to the upper range limit. It’s thus advis-
able to perform measurement with a fixed measuring range. Due
to the fact that rapid momentary value acquisition (see section
4.5.1) has a large measuring range span, limit values of greater
than 30,000 digits can be selected.
Actual measurement is always executed using the sampling rate
selected in “Store > rAtE”.
1nFo
... store ... trig
trig set : sto-ou / sto-in / off
l_trig set : -150000 ... +150000
h_trig set : -150000 ... +150000
tStorE – Recording Time (parameter for memory mode operation)
This parameter determines whether or not measured values will
be recorded for a limited time only. If recording time will be limited,
its duration can be entered in hours, minutes and seconds. “On”
means unlimited recording time.
1nFo
... store ... t .store
t .store set : on /99:00:00 (hh:mm:ss)