User Manual

46 GMC-I Messtechnik GmbH
V/Hz, , Temperature, and A/Hz Measurements
5.7.4 Alternating Current Measurement with Current Clamp Sensor,
A AC and Hz
Voltage/Current Transformer Output
When a current clamp sensor is connected to the multimeter (V
input), all current displays appear with the correct value in accor-
dance with the selected transformation ratio. The only prerequisite is
that the current sensor is equipped with at least one of the below
listed transformation ratios, and that the ratio has been previously
selected in the following menu (
Cl
ip

0FF
) (see also section 6.4).
Current Clamp Setup Menu
1nFo
... SET t iME ... CLIP
CLIP
set:1 / 10/100/1000 / 0ff 
The maximum allowable operating voltage is equal to the nominal
voltage of the current transformer.
When reading the measured
value, additional error resulting from the current clamp sensor must
also be taken into consideration
(default setting: 1:1000).
Transforma-
tion ratios:
CL IP
DMM Measuring Ranges Clamp Types
300 mV 3 V 30 V
1:1
1mV/1mA
300.00 mA 3.0000 A 30.000 A WZ12C
1:10
1mV/10mA
3.0000 A 30.000 A 300.00 A
WZ12B, Z201A/B
METRAFLEX
1:100
1mV/100mA
30.000 A 300.00 A 3000.0 A
Z202A/B,
METRAFLEX
1:1000
1 mV/1 A
300.00 A 3000.0 A 30000 A
Z202A
/B
, Z203A
/B
,
WZ12C,
METRAFLEX
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
003.50
A
AC
TRMS
A~
Hz
1:1000
050.00
Hz
AC
1:1000
FUNC
ENTER
COM
V
Tem p
A
~
R
x
Current
Ri
Ri = 5 M
mA
050
.
00
A
015
.
000
Hz
V~
Hz
R
i
= 5 M