V/Hz, , Temperature, and A/Hz Measurements
5.4 Diode Testing with Constant Current of 1 mA
Ð Disconnect supply power from the electrical circuit of the de-
vice to be measured, and discharge all high-voltage capaci-
tors.
Ð Make sure that the device under test is voltage-free. Interfer-
ence voltages distort measurement results!
Refer to section 5.1.1 regarding testing for the absence of volt-
age with the help of the direct voltage measurement.
Ð Set the rotary switch to .
Ð Acknowledge by pressing the
FUNC
| ENTER
key.
Ð Connect the device under test as shown.
Forward Direction and Short-Circuit
The instrument displays forward voltage in volts (display: 4
places). As long as voltage drop does not exceed the maximum
display value of 4.5 V, several series connected components or
reference diodes can be tested with a small reference voltage and
Z-diodes.
Reverse Direction and Interruption
The measuring instrument indicates overload:0 L”.
Note
Resistors and semiconductor paths connected in parallel to
the diode distort measurement results!
Attention!
!
Observe high open-circuit voltage of 6 V during diode test-
ing: Circuits must be laid out accordingly.