Ending Recording
Ð If the instrument is in the measuring mode, return to the menu
function by pressing the MEASURE | SETUP key. Select “StorE
again and acknowledge by pressing the FUNC | ENTER key.
StoP” blinks at the main display.
store on StoP store
Ð Acknowledge the “StoPdisplay by pressing FUNC | ENTER.
The
STORE display segment in the header is cleared, indicating
that recording has been ended.
Ð Press the MEASURE | SETUP key in order to return to the measur-
ing function.
Ð Memory mode operation can also be exited by switching the
multimeter off.
Querying Memory Occupancy
Memory occupancy can be queried during recording with the
help of the “1nFo” menu (see also section 6.3).
Memory occupancy range: 000.1% to 099.9%.
1nFo
bAtt
...
0CCvP
%
017
.
4
%
Clear Memory
This function deletes all measured values from memory!
It’s advisable to execute this function before starting a new mea-
surement data recording.
1nFo
... StorE store off StArt (blinks)
store CLEAr store CLEAr no yes
store empty (appears briefly) store