Datasheet
GMC-I Messtechnik GmbH
5
METRAHITULTRA
High Resolution TRMS System Multimeter
Internal Measured Value Storage
Memory capacity 16 MBit for approx. 300,000 measured
values with indication of date and time
Power Supply
Battery 2 ea. 1.5 V mignon cell (2 ea. size AA)
Alkaline manganese per IEC LR6
(2 ea. 1.2 V NiMH rechargeable battery
also possible)
Service life With alkaline manganese: approx. 200 hrs.
Battery indicator Battery capacity display with battery sym-
bol in 4 segments: .
Querying of momentary battery voltage via
menu function.
Power OFF function The multimeter is switched off automati-
cally:
– If battery voltage drops to below
approx. 1.8 V
– If none of the keys or the rotary switch
are activated for an adjustable duration
(10 to 59 min.) and the multimeter is not
in the continuous operation mode
Power pack socket
If the NA X-TRA power pack has been
plugged into the instrument, the installed
batteries are disconnected automatically.
Rechargeable batteries can only be
recharged externally.
Display
Transreflective
LCD panel (65
mm x 36 mm)
with display of up
to 3 measured
values, unit of
measure, type of
current and vari-
ous special functions.
Background illumination
Background illumination is switched off approximately 1 minute
after it has been activated.
Digital
Display / char. height 7-segment characters
Main display: 13 mm
Auxiliary display: 7.5 mm
Number of places
309,999
steps
Overflow display “OL” is displayed for 310,000 digits
Polarity display “–” (minus sign) is displayed if plus pole is
connected to “”
Measuring rate 10 or 40 measurements per second
with
the Min-Max function except for the capaci-
tance and frequency measuring functions
Refresh Rate 2 times per second, every 500 ms
Acoustic Signals
For voltage Above 600 V in the 600 V range: intermit-
tent (250 ms on, 250 ms off)
For current – Above 10 A: intermittent signal
– Above 16 A: continuous signal
– For displayed temperature > 50 °C
Fuse
Fuse link
FF (UR) 10 A/1000 V AC/DC, 10 x 38 mm,
Switching capacity: 30 kA at 1000 V AC/DC,
protects the current measurement input in
the 300 A through 10 A ranges
Electrical Safety
per IEC 61010-1:2010/VDE 0411-1:2011
Protection class
II
Measuring category CAT III CAT IV
Operating voltage 600 V 300 V
Pollution degree
2
Tes t Voltag e
5.2 kV~
Electromagnetic Compatibility (EMC)
Interference emission EN 610326-1:2006 class B
Interference immunity EN 610326-1:2006
EN 610326-2-1:2006
Ambient Conditions
Accuracy range 0 C to +40 C
Operating temperature
range T
A
–100 C to +50 C *
Storage temp. range –25 C to +70 C (without batteries)
Relative humidity 40 to 75%, no condensation allowed
Elevation To 2000 m
Deployment Indoors, except within specified ambient
conditions
* Exception: current > 10 A to 16 A, operation at up to 40 °C
Mechanical Design
Housing Impact resistant plastic (ABS)
Dimensions 200 x 87 x 45 mm (without rubber holster)
Weight Approx. 0.4 kg with batteries
Protection Housing: IP 52 (pressure equalization by
means of the housing)
Sockets: IP 20
Table Excerpt Regarding Significance of IP
Codes
IP XY
(1
st
digit X)
Protection Against
Foreign Object Entry
IP XY
(2
nd
digit Y)
Protection Against
Penetration by Water
2 12.5 mm dia. 0 Not protected