Datasheet
GMC-I Messtechnik GmbH
3
METRAHITULTRA
High Resolution TRMS System Multimeter
Characteristic Values
1
Specified accuracy valid as of 1% of the measuring range
2
Specified accuracy valid as of 2% of the measuring range
3
Display of up to max. 4.5 V, above which overload display appears: “OL”
4
Applies to measurements at film capacitors during battery operation
5
Lowest measurable frequency for sinusoidal measuring signals symmetrical to the zero point
6
Overload capacity of the voltage measurement input:
power limiting: frequency x voltage max. 3 x 10
6
V x Hz for U > 100 V
7
Overload capacity of the current measurement input:
see current measuring ranges
for max. current values.
8
Input sensitivity, sinusoidal signal: 10% to100% of voltage/current
measuring range
;
in the 300 kHz range the specified intrinsic uncertainty applies as from 15% of MR
9
Plus sensor deviation
10
As of a measured value of 7 A, measurement is limited to an ambient temperature
of 30 C or a maximum duration of 5 minutes.
11
Off-time > 30 min. and T
A
40 C after a 10 or 16 A measurement
12
At 0 ... + 40 C
Key: d = digit(s), MR = measuring range, rdg. = reading (measured value)
Meas.
Function
Measuring Range
Resolution at Upper Range Limit
Input Impedance
Intrinsic Uncertainty under Reference Conditions
Overload
Capacity
12
(...% rdg + % MR +... d)
(... % rdg. + ... d)
(... % rdg. + ... d)
DC AC/AC+DC
309,999 30,999 3099 ~ / ~
2
Value Time
V
300 mV 1 V10V 10 M > 5 M // < 50 pF
0.02 + 0.005 + 10 with ZERO
0.5 + 30
2
0.5 + 30
600V
DC
AC
RMS
Sinusoi-
dal
Max. 10
sec.
3V 10V 100 V 10 M > 5 M // < 50 pF 0.02 + 0.005+ 5
0.2 + 30
1
0.5 + 30
Cont.
30 V 100 V1mV 10 M > 5 M // < 50 pF 0.02 + 0.005 + 5
300 V 1mV 10mV 10 M > 5 M // < 50 pF 0.02 + 0.005 + 5
600 V 10 mV 100 mV 10 M > 5 M // < 50 pF 0.02 + 0.005 + 5 0.2 + 30 0.5 + 30
Display Range for Reference Voltage
U
REF = 0.775 V
Intrinsic
Uncertainty
dB
0.3 V / 3 V ... 600 V~ 0.01 dB -42 dB ... + 57 dB
0.1 dB (U > 10% MR)
600 V
AC RMS
sine
Cont.
DC AC/AC+DC
Voltage drop at approx. range limit ~
2
2
A
300 A 1 nA 10 nA 65 mV 0.05 + 0.02 + 5 with ZERO
0.5 + 30 0.5 + 30 0.7 A Cont.
3 mA 10 nA 100 nA 170 mV 0.05 + 0.01 + 5
30 mA 100 nA 1 A 170 mV 0.02 + 0.01 + 5
300 mA 1 A10A 200 mV 0.1 + 0.05 + 5
3A 10A 100 A 150 mV 0.2 + 0.05 + 5 with ZERO 0.7 + 30 0.7 + 30
10 A:
5 min.
10, 11
16 A: 30 s
11
10 A 100 A 1 mA 470 mV 0.2 + 0.05 + 5 0.5 + 30 0.5 + 30
Factor: 1:1/10/100/1000
Input Input impedance
A
0.03, 0.3, 3, 30 A 30 mA
Current measurement input
(A socket)
See current measuring ranges for specification.
Meas. input
0.3, 3, 30, 300 A 300 mA 0.7 A cont.
3A: 5 min.
3/30/300/3000 A
3 A Plus current transformer clamp error
A
0.3, 3, 30, 300 A
3/30/300/3000 A
300 mV
3V / 30V
Voltage measurement input
(V jack) Ri =5 M
/10 M
See voltage measuring ranges for specification.
Meas. input
600 V TRMS
Open-circuit
voltage
Measuring current at
range limit
(... % rdg. + % MR ... d)
300 1m < 2 V Approx. 0.5 mA
0.05 + 0.01 +5 with ZERO function active
600 V
DC
AC
RMS
Sinusoi-
dal
Max. 10
sec.
(PTC)
3k 10 m < 2 V Approx. 130 A 0.05 + 0.01 + 5
with ZERO function active
30 k 100 m < 2 V Approx. 20 A 0.05 + 0.01 + 5
300 k 1 < 2 V Approx. 2 A 0.05 + 0.01 + 5
3M 10 < 2 V Approx. 1 A 0.1 + 0.02 + 5
30 M 100 < 2 V Approx. 200 nA 1 + 0.2 + 5
300 —0.1 < 4.5 V Approx. 1 mA const. 1 + 5
with ZERO function
active
600 V
Max. 10
sec.
4,5 V
3
— 1 mV < 6 V Approx. 1 mA const. 0.2 + 3 600 V
Max. 10
sec.
Discharge
resistance
U
0 max
(... % rdg. + ... d)
4
F
3nF — — 1pF 1M 2 V 2 + 15
with ZERO function active
600 V
DC
AC
RMS
sine
Max.
10 s
30 nF — — 10 pF 1 M 2V 1 + 6
with ZERO function active
300 nF — — 100 pF 100 k 2V
1 + 63 F — — 1 nF 100 k 2V
30 F— — 10nF 10k 2V
300 F — — 100 nF
2.5 k
2V
5 + 6
3mF — — 1μF 2 V
f
min
5
(... % rdg. + ... d)
Hz (V)
Hz (A)
Hz (A )
300 Hz 0.001Hz
5 Hz Hz(V) 0.05 + 2
8
Hz(A) 0.05 + 3
8
Hz (V)
6
:
Hz(A
6
600 V
Hz (A):
7
Max.
10 sec.
3 kHz 0.01 Hz
30 kHz 0.1 Hz
Hz (V)
300 kHz 1 Hz 10 Hz
MHz
300 Hz to 1 MHz
0.01 ...100 Hz
1 Hz > 3 V ... 5 V
600 V
Max. 10
sec.
%
2.00 to 98.00% — 0.01% 15 Hz ... 1 kHz 0.1 MR
10 d > 3 V ... 5 V
600 V
Max.
10 sec.
5.00 to 95.00% — 0.01% 1 ... 10 kHz
0.1 MR per kHz
10 d
> 3 V ... 5 V
(... % rdg. + ... d)
C/F
Pt100
Pt1000
–200.0 ...
+100.0 C
0.1 K
0.3 + 10
9
600 V
DC/AC
RMS
sine
Max.
10 sec.
+ 100.0 ...
+850.0 C
K
(NiCr-Ni)
–270.0 ...
+1372.0
C
1% + 2.0 K
9
Internal
temp.
mea-
sure-
ment
–10 ...
+80 C
0.1 K Auxiliary display in ampere range 2 K