User Manual
36 GMC-I Gossen-Metrawatt GmbH
V/Hz, Ω, Temperature,   and A/Hz Measurements
5.4 Diode Testing   with a Constant Current of 1 mA
➭ Disconnect supply power from the electrical circuit of the 
device to be measured, and discharge all high-voltage 
capacitors.
➭ Make sure that the device under test is voltage-free. 
Interference voltages distort measurement results!
Refer to chapter 5.1.1 regarding testing for the absence of 
voltage with the help of the direct voltage measurement.
➭ Set the rotary switch to  .
➭ Press the 
FUNC
 | ENTER
 key.
➭ Connect the device under test as shown.
Conducting Direction and Short-Circuit
The instrument displays forward voltage in volts (display: 4 
places). As long as voltage drop does not exceed the maximum 
display value of 5.1 V, several series connected components or 
reference diodes can be tested with a small reference voltage and 
reference diodes.
Reverse Direction and Interruption
The measuring instrument indicates overload 0L. 
Note
Resistors and semiconductor paths connected in parallel to 
the diode distort measurement results!
COM
V
Temp
Ω
Forward Direction Reverse Direction
0V !
0.654
0.L
to 5.100 V
Measuring Range:
+
–
0
.
443
V
!
000
.
8
Ω
DC
FUNC
ENTER