Operation Manual

GMC-I Messtechnik GmbH 41
Measurements: V/Hz – R
SL
Ω
– Temp – – m – M
Ω
ISO
– A/Hz
Ð Starting the insulation resistance measurement:
Briefly press the U
ISO ON / OFF key.
Insulation resistance is displayed for the currently selected
conductor pair.
Ð Reversing polarity of the cable under test:
Normal polarity reversal:
Press and hold the POL / U
ISO key.
Rapid polarity reversal:
Press the POL / UISO key at short intervals. bAL.C appears at
the display (ballistic capacitance) for relative cable length
determination. After key activation has ceased for a period
of approximately 2 seconds, the instrument is switched
back to standard insulation resistance measurement.
Ð Switch the rotary selector switch to the MΩ
ISO
_a-b, MΩ
ISO
_a-
E and MΩ
ISO
_b-E positions, one after the other, in order to
execute the desired tests.
Auto-ranging is active during insulation resistance measurement.
If the measured value is less than 10% of the measuring range
after manual measuring range selection (see chapter 4.1.2), u r
(under-range) appears at the display. You should then select the
next smaller measuring range with the help of the
key.
Automatic Recognition of Interference Voltage During
Insulation Resistance Measurement
If, during insulation measurement, the instrument detects an
interference voltage of greater than 15 V AC or 25 V DC (assuming: U
in-
terference
U
ISO
, Riq < 100 kΩ), „Error“ is briefly displayed at the
LCD. The instrument is then automatically switched to voltage
measurement, and the currently measured voltage value is dis-
played.
000.00
M
Ω
MΩ
a-b
a-E
b-E
MΩ
iso
at 100 V
iso
a-b
FUNC
ENTER
UISO ON / OFF
ZERO
ESC
POL / UISO
FUNC
ENTER
UISO ON / OFF
a-E
FUNC
ENTER
UISO ON / OFF
ZERO
ESC
POL / UISO
FUNC
ENTER
UISO ON / OFF
b-E
FUNC
ENTER
UISO ON / OFF
ZERO
ESC
POL / UISO
FUNC
ENTER
UISO ON / OFF
UISO
000.0
UISO
V
FUNC
ENTER
+–
+
a
a
b
b
E
E
+
0V !
Reverse polarity
Discharge
Switch test
voltage on
DC
AC
TRMS
100 V
iso