Operation Manual
GMC-I Messtechnik GmbH 67
Technical Data
Power Supply
Battery 2 ea. 1.5 V mignon cell (2 ea. size AA),
alkaline manganese per IEC LR6
Service life With alkaline manganese batteries: approx. 
200 hours without MΩ
ISO
 measurement
Battery test Battery capacity display with battery 
symbol in 4 segments:  . 
Querying of momentary battery voltage via 
menu function.
Power OFF function
The multimeter is switched off automatically: 
– If battery voltage drops to below approx.
1.8 V
– If none of the keys or the rotary switch are
activated for an adjustable duration of 10
to 59 minutes, and the multimeter is not in
the continuous operation mode 
Power pack socket
If the power pack has been plugged into 
the instrument, the installed batteries are 
disconnected automatically. 
Rechargeable batteries can only be 
recharged externally.
1)
Times 0.7 for interface operation
Electromagnetic Compatibility (EMC)
Interference emission EN 61326:2006-1 class B
Interference immunity EN 61326:2006-1
EN 61326:2006-2-1
Data Interface
Type
Optical via infrared light through the housing
Data transmission Serial, bidirectional (not IrDa compatible)
Protocol Device specific
Baud Rate 38,400 baud
Functions –  Select/query measuring functions 
and parameters
–
Query momentary measurement data
The USB⏐X-TRA plug-in interface adapter (see accessories) is 
used for adaptation to the PC’s USB port.
Internal Measured Value Storage
Memory capacity 4 MBit / 540 kB for approx. 15,400 
measured values with indication of date 
and time
Measuring 
Function
Nominal 
Voltage U
N
Resistance
of the 
DUT
Service life 
in Hours
Number of Possible 
Measurements with 
Nominal Current per 
VDE 0413
V 200 
1)
V 150 
1)
MΩ
10 V/100 V 1 MΩ 50
10 V/100 V 100 kΩ 3000