Operation Manual

GMC-I Messtechnik GmbH 63
Technical Data
1)
15 ... 45 ... 65 Hz ... 10 (5) kHz sine. See following page for influences.
2)
At 0° ... + 40° C
3)
Display of up to max. 5.1 V, “OL” in excess of 5.1 V.
4)
Applies to measurements at film capacitors and with battery operation
5)
Lowest measurable freq. for sinusoidal meas. signals symmetrical to zero point
6)
Overload capacity of the voltage measurement input: power limiting:
frequency x max. voltage 6 x 10
6
V x Hz: e.g. max. 600 V 1 kHz
7)
Overload capacity of the current measurement input:
See current measuring ranges for maximum current values.
8)
Input sensitivity, sinusoidal signal, 10% to 100% of the voltage or current measuring
range; limitation: up to 30% of the range at up to 100 kHz in the mV measuring range,
50% of the range at up to 300 kHz, 30% of the range in the 1 A measuring range.
The voltage measuring ranges with max. 30 kHz apply in the A measuring range.
9)
Plus sensor deviation
10)
With ZERO function active
11)
Residual value of 1 to 10 d with short circuited terminal tips,
exception: mV /
μ
A range of 1 to 35 d at zero point due to the TRMS converter
Key: R = meas. range, d = digit(s), rdg. = measured value (reading)
Discharge resist. U
0 max
± (... % rdg. + ... d)
F
(a)
30.00 nF 10 pF 10 MΩ 0.7 V 1 + 6
4), 10)
600 V
DC
AC
RMS
Sine
Max. 10 s
300.0 nF 100 pF 1 MΩ 0.7 V 1 + 6
4)
3.000 μF 1 nF 100 kΩ 0.7 V 1 + 6
4)
30.00 μF10nF 12kΩ 0.7 V 1 + 6
4)
300.0 μF 100 nF 3 kΩ 0.7 V 5 + 6
4)
f
min
5)
± (... % rdg. + ... d)
Hz (V)/
Hz (A)
300.0 Hz 0.1 Hz
1Hz
0.5 + 1
8)
Hz (V)
6)
:
Hz(A )
6)
:
600 V
Hz (A):
7)
Max. 10 s
3.000 kHz 1 Hz
30.00 kHz 10 Hz 10 Hz
Hz (V)
300.0 kHz 100 Hz 100 Hz
± (... % rdg. + ... d)
9)
°C
Pt100
– 200.0 ...
+200.0 °C
0.1 °C
2 K + 5
600 V
DC/AC
RMS
Sine
Max. 10 s
+200.0 ...
+850.0 °C
1 + 5
Pt1000
– 150.0 ...
+200.0 °C
2 K + 5
+200.0 ...
+850.0 °C
1 + 5