Device and Measuring Parameters
A.diSP – analog display: select display mode
One of two different display modes can be selected for the analog 
display:
• bArG: bar graph
• Po int: pointer
 1nFo 
 ...  SET rAtE  ...  A.d iSP
 bArG / Po int 
 
(po int = default setting)
APoFF – specified time for automatic shutdown and continuous ON
The instrument is switched off automatically if the measured value 
remains unchanged for a long period of time and if none of the 
keys or the rotary switch have been activated before the specified 
time “APoFF” (entered in minutes) has elapsed.
If the on setting is selected, the multimeter is set for long-term 
measurement and 
ON appears in the display to the right of the 
battery symbol. In this case, the multimeter can only be switched 
off manually. The “on” setting can only be cancelled by changing 
the respective parameter, and not by switching the instrument off.
 1nFo 
 ...  SET rAtE  ...  APoFF
 10 ... 59 min on 
   
(10 minutes = default setting)
U_ISo – select test voltage
The desired test voltage for insulation resistance measurement 
can be selected here:
 1nFo 
 ...  SET rAtE  ...  v1so
 ... 10, 100 V 
   
bEEP – set limit value for continuity testing
 1nFo 
 ...  SET rAtE  ...  bEEP
 1, 10, 20, 30, 40, 90 Ω 
 
(10 Ω = default setting)
irStb – status of the infrared receiver in the stand-by mode
See chapter 8.2 on page 61 regarding settings.
Addr – set device address
See chapter 8.2 on page 61.