Operation Manual
52 GMC-I Messtechnik GmbH
Measurements R
SL
 – MΩ
ISO
@10 V/100 V – CAP Switch positions (a-b) (a-E) (b-E)
NTBA connected
If an NTBA is connected, measurement is not possible as the 
input circuit of the NTBA prevents recognition 
(influence value = –OL)
DSL modem active
If a DSL modem is active and the connector cable is very short 
(< 500 m), an SM100 signature adapter must be plugged on in 
oder to attenuate the DSL signals (handshake signals). The 
SM100 adapter attenuates the acoustic handshake signals, thus 
making measurement possible.
This adapter, however, must be removed from the instrument for 
all other tests and measurements except for the capacitance 
measurement with connected DSL modem/router described 
above, see chapter 6.4.