Operation Manual
48 GMC-I Messtechnik GmbH
Measurements R
SL
 – MΩ
ISO
@10 V/100 V – CAP Switch positions (a-b) (a-E) (b-E)
Auto-ranging is active during insulation resistance measurement. 
If the measured value is less than 10% of the measuring range 
after manual measuring range selection (see chapter 4.1.2), u r 
(under-range) appears at the display. You should then select the 
next smaller measuring range with the help of the 
 key.
Automatic Recognition of Interference Voltage During 
Insulation Resistance Measurement (only in AUTORANGE Mode)
If, during insulation measurement, the instrument detects an 
interference voltage of greater than 15 V AC or 25 V DC (assuming: U
in-
terference
 ≠ U
ISO
, Riq < 100 kΩ), „Error“ is briefly displayed at the 
LCD. The instrument is then automatically switched to voltage 
measurement, and the currently measured voltage value is dis-
played.
Note
A polarity-dependent dead zone results in erroneous mea-
surements for automatic interference voltage detection. The 
dead zone lies within a range of 60 to 135 V DC AC
TRMS
sinusoidal (in the case of an interference voltage whose 
value is equal to that of measuring voltage, the two voltages 
neutralize each other).
Manual switching to insulation resistance measurement is 
disabled for as long as voltage is applied to the test terminals. 
If interference voltage is no longer present, the MΩ
ISO
measurement can be started by once again pressing the 
UISO ON / OFF key.
If a voltage of greater than 110 V DC AC is present, it is indicated by 
means of an acoustic signal, as well as optically with the “UHI” 
symbol at the display. 
Attention!
!
If „Error“ appears at the display, the cable (the device under 
test) is most likely capacitively charged to a 
significant extent. Remedy: Short circuit conductors 
a-b, a-E and b-E. Repeat the measurement.
6.3.4 Ending the Measurement and Discharging
Ð Briefly press the UISO ON / OFF key.
After measurement has been completed, any remaining residual 
voltage is displayed which may result from cable capacitance. 
The instrument’s internal 120 kΩ resistor causes rapid 
discharging. However, contact to the device under test must be 
maintained. The falling voltage value can be observed directly at 
the LCD. Do not disconnect the device under test until the voltage value 
has dropped to below 25 V!
Note
The instrument’s batteries are rapidly depleted during insulation 
resistance measurement. Deactivate insulation resistance mea-
surement between measurements for this reason. Use only 
alkaline manganese batteries in accordance with IEC 6 LR61.
Note
Rotary selector switch positions a-b, a-E and b-E are 
intended solely for the detection of interference voltage up to 
110 V during insulation resistance measurement.
Voltage measurement may only be performed with the rotary 
selector switch set to the V , V  or the V  position. 
The DATA, MIN/MAX and ZERO functions are not available 
in this case.