Operation Manual
46 GMC-I Messtechnik GmbH
Measurements R
SL
 – MΩ
ISO
@10 V/100 V – CAP Switch positions (a-b) (a-E) (b-E)
6.3.1 Connecting the Measurement Cables
Note
Testing the Measurement Cables 
The test probes at the ends of the measurement cables 
should be short circuited before performing insulation 
resistance measurements with the selector switch in the Ω 
or the   position, in order to make sure that a value close 
to 0 Ω is displayed at the instrument. Incorrect connection or 
a broken measurement cable can be detected in this way.
Ð Cable connection: 
Connect the measurement cables to jacks a, b and E.
6.3.2 Selection of Test Voltage or Galvanic Signature
In the interference voltage detection mode you can select test 
voltages of 10 V or 100 V with the 
 or  key. The selected test 
voltage is shown as a segment above the digital display, the sig-
nature recognition mode is shown as a digital display.