Operation Manual
GMC-I Messtechnik GmbH 41
Measurements: V/Hz – 
Ω
 –   – Temp –   – m – A/Hz
5.7.3 Direct and Pulsating Current Measurement 
with Clip-On Current Sensor, A DC and A (DC+AC)
Transformer Output, Voltage/Current
When a clip-on current sensor is connected to the multimeter 
(V input), all current displays appear with the correct value in 
accordance with the selected transformation ratio. The only 
prerequisite is that the current sensor is equipped with at least 
one of the below listed transformation ratios, and that the ratio 
has been previously selected in the following menu (Cl ip ≠ 0FF) 
(see also chapter 7.4).
Current Clip Setup Menu
 1nFo 
 ...  SET   rAtE  ...  CL IP
 0ff/1 / 10 /100/1000 
 
Maximum allowable operating voltage is equal to the current 
transformer’s nominal voltage. 
When reading the measured value, 
additional error resulting from the clip-on current sensor must also be 
taken into consideration (
default value: Cl ip = 0ff = voltage dis-
play).
Trans. Ratio
CL IP
Measuring Ranges Clip Type
300 mV 3 V 30 V
1:1 
1 mV / 1 mA
300.0 mA 3.000 A 30.00 A WZ12C
1:10 
1m V / 10 mA
3.000 A 30.00 A 300.0 A
WZ12B, Z201A/B, 
METRAFLEX
1:100 
1m V / 100 mA
30.00 A 300.0 A 3,000 kA
Z202A/B, 
METRAFLEX
1:1000 
1 mV/1 A
300.0 A 3,000 kA 30.00 kA
Z202A/B, Z203A/B, 
WZ12C, 
METRAFLEX
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Tem p
Ω
aEb
~
R
x
Current
0035
.
0
A
1:1000
A
DC
0035
.
0
A
1:1000
A
DC
AC
TRMS
FUNC
ENTER
R
i
R
i
 ~ 9 MΩ
V
V
MEASURE
SETUP
 ...  SET  ...  CL IP
FUNC
ENTER
FUNC
ENTER
  0ff
Measuring ranges:
mA / A: see table