Operation Manual
32 GMC-I Messtechnik GmbH
Measurements: V/Hz – 
Ω
 –   – Temp –   – m – A/Hz
5.3 Temperature Measurement – Temp RTD
Temperature measurement is performed with a Pt100 or Pt1000 
resistance thermometer (accessory, not included), which is con-
nected to the voltage input.
Selecting the Unit of Measure for Temperature
 1nFo 
 ...  tEMP   °C / °F  
(°C = default setting)
5.3.1 Measurement with Resistance Thermometers
Ð Set the rotary switch to “Ω” or “Temp
RTD
”.
Press the FUNC | ENTER key in order to change to the other measuring func-
tion if required.
The sensor type, i.e. Pt100 or Pt1000, is detected automatically and 
displayed.
There are two different ways to compensate for cable resistance:
Automatic Compensation
Ð Acknowledge by pressing the ZERO | ESC key.
“Short leads” appears at the display.
If you prefer to enter cable resistance directly, you can skip the 
following entry prompt.
Ð Short circuit the measuring instrument’s connector cables. 
“000.0” appears at the display. After pressing the 
FUNC | ENTER key, automatic compensation of cable 
resistance is activated for all subsequent measurements. 
The short-circuit can now be eliminated, and the device is 
ready for use.
Enter Cable Resistance
Ð Press the ZERO | ESC key once again in the automatic 
compensation menu.
Ð Enter the known resistance of the connector cables with the 
scroll keys: 
Select the digit to be changes with the 
  keys, and change 
the respectively selected digit with the  keys. The default 
value is 0.43 Ω. Values can be selected within a range of 
0 to 50 Ω.
Ð Upon pressing the FUNC | ENTER key, the selected value is acti-
vated and the display is returned to the measuring function. 
Cable resistance remains in memory even after the instrument 
has been switched off.
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER