5.1.3 Transient Overvoltages
The multimeters are protected against transient overvoltages of 
up to 4 kV with wave-front durations of 1.2 ms and halftimes of 
50 μs in the voltage measuring range. If longer pulse durations are 
expected, for example when conducting measurements at trans-
formers or motors, we recommend the use of our KS30 measur-
ing adapter. It provides protection against transient 
overvoltages of up to 6 kV with wave-front durations of 10, and 
halftimes of 1000 μs. Continuous load capacity amounts to 
1200 V
RMS
. Additional influence error caused by the KS30 
measuring adapter amounts to approximately -2%.