Operation Manual
28 GMC-I Messtechnik GmbH
Measurements: V/Hz – 
Ω
 –   – Temp –   – m – A/Hz
5.1.2 Alternating Voltage and Frequency Measurement V AC and Hz 
with Selectable Low-Pass Filter
Ð Set the CL iP parameter to 0FF in the current clip setup menu. 
Otherwise all measured values are displayed in amperes, and 
are corrected by the amount resulting from the selected trans-
formation ratio for an interconnected clip-on current sensor.
 1nFo  ...  SET   rAtE  ...  CL IP 
 1 / 10/100/1000 / 0ff 
Ð In accordance with the voltage or frequency to be measured, 
turn the rotary switch to V~ or Hz.
Ð Connect the measurement cables as shown. 
The “⊥” connector jack should be grounded.
Voltage Measurement
Note
An intermittent acoustic signal warns the operator if the 
measured value exceeds the upper range limit in the 600 V 
range.
Make sure that a current measuring range (“A”) has not been 
activated, when the multimeter is connected for voltage 
measurement! If the fuse’s blowing limits are exceeded as a result 
of operator error, both the operator and the instrument are in 
danger!
Ð You can switch back and forth between voltage measurement 
with and without low-pass filter.
Ð Repeatedly press the FUNC | ENTER multifunction key, until the V 
or V/Fil unit of measure appears at the display.
Frequency Measurement
Ð Apply the measured quantity is the same way as for voltage 
measurement.
Ð Manually select the measuring range for the voltage amplitude. 
When the instrument is switched to frequency measurement, 
the previously selected voltage measuring range remains ac-
tive.
Ð You can switch back and forth between frequency 
measurement with and without low-pass filter.
Repeatedly press the FUNC | ENTER multifunction key, until the 
Hz or Hz/Fil unit of measure appears at the display. 
Lowest measurable frequencies and maximum allowable 
voltages are included in chapter 9, “Technical Data”.
Measurement with Low-Pass Filter
Attention!
!
Be aware of the fact that dangerous voltage spikes are not 
displayed during this type of measurement (see also 
“Voltage Comparator”.
We recommend measuring voltage without the low-pass 
filter first, in order to be able to detect any dangerous 
voltages.
A 200 Hz low-pass filter can be activated if required, in order to 
filter out capacitively induced high frequency pulses of greater 
than 200 Hz, for example when performing measurements at 
cables, i.e. undesired voltages of greater than 200 Hz can be 
suppressed.