Measurements: V/Hz –
Ω
– Temp m – A/Hz
5.1.1 Direct and Pulsating Voltage Measurement, V DC and V (DC+AC)
Ð Set the CL iP parameter to 0FF in the current clip setup menu.
Otherwise all measured values are displayed in amperes, and
are corrected by the amount resulting from the selected trans-
formation ratio for an interconnected clip-on current sensor.
1nFo
... SET rAtE ... CL IP
1 / 10/100/1000 / 0ff

Ð In accordance with the voltage to be measured, turn the rotary
switch to V or V .
Ð Connect the measurement cables as shown.
The “” connector jack should be grounded.
Note
An intermittent acoustic signal warns the operator if the
measured value exceeds the upper range limit in the 600 V
range.
Make sure that a current measuring range (“A”) has not been
activated when the multimeter is connected for voltage
measurement! If the fuse’s blowing limits are exceeded as a result
of operator error, both the operator and the instrument are in
danger!
With the rotary switch in the V position, the multimeter is always in
the 3 V measuring range immediately after it is switched on. As
soon a the MAN / AUTO key is pressed, and assuming the
measured value is less than 280 mV, the multimeter is switched to
the mV measuring range.
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Temp
Ω
aEb
– (+)
+ (–)
V= : 100 μV610 V
020.00
V
DC
AC
TRMS
020.00
V
DC
V
V
V
V
> 15 V AC or > 25 V DC:
> 620 V:
Warnings regarding dangerous voltage:
Max. 600 V 1 kHz
V: 10 mV610 V
Hz: 1Hz 300 kHz
P
max
= 6 x 10
6
V x Hz
Measuring ranges:
230
.
0
FUNC
ENTER
MEASURE
SETUP
... SET ... CL IP
FUNC
ENTER
FUNC
ENTER
 0ff
5 ranges: 300 mV, 3 V, 30 V,
300 V and 600 V