Operation Manual
24 GMC-I Messtechnik GmbH
Control Functions
During Recording
STORE is displayed underneath the analog display during memory 
mode operation, and 
memory occupancy
 can be monitored:
StoP 
 000.3%
The following message appears as soon as memory is full: 
“100.0%”.
In order to be able to observe measured values during recording, 
switch to the measuring function by pressing MEASURE | SETUP. 
The display is returned to the memory menu after once again 
pressing MEASURE | SETUP.
A new memory block is created when another measuring function 
is selected with the rotary switch or the FUNC | ENTER key. Data 
storage then continues automatically.
Ending Recording
Ð “StoP” appears at the display after pressing MEASURE | SETUP.
StoP start
Ð Acknowledge the “StoP” display by pressing FUNC | ENTER. 
STORE is cleared from the display indicating that recording has 
been ended.
Ð Press MEASURE | SETUP in order to return to the measuring 
function.
Ð Memory mode operation can also be exited by switching the 
multimeter off.
Querying Memory Occupancy
Memory occupancy can be queried during recording with the 
help of the “1nFo” menu (see also chapter 7.3).
Memory occupancy range: 000.1% to 099.9%.
1nFo
bAtt
: 
 ... 
0CCUP
 %: 
017.4
 %
Memory occupancy can be queried before recording is started via 
the “StorE” menu.
 1nFo 
 ...  StorE 017.4 %  StArt 
Clear Memory
This function deletes all measured values from memory!
This function cannot be executed during memory mode 
operation.
 1nFo 
 ...  StorE   017.4 %  StArt 
 CLEAr   empty
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER