Operation Manual
18 GMC-I Messtechnik GmbH
Control Functions
4 Control Functions
4.1 Selecting Measuring Functions and Measuring Ranges
4.1.1 Automatic Range Selection
The multimeter is equipped with auto-ranging for all measuring
functions, except for temperature measurement, and diode and
continuity testing. Auto-ranging is active as soon as the
instrument is switched on. The instrument automatically selects
the measuring range which allows for highest possible resolution
of the applied quantity. When the instrument is switched to
frequency measurement, the previously selected voltage
measuring range remains active.
AUTO-Range Function
The multimeter is switched automatically to the next higher range
at ±(3099 d + 1 d → 0310 d), and to the next lower range at
±(280 d - 1 d → 2799 d).
4.1.2 Manual Measuring Range Selection
Auto-ranging can be deactivated and measuring ranges can be
selected manually in accordance with the following table by
pressing the MAN / AUTO button.
The desired measuring range can then be selected with the
or
key.
The instrument is automatically returned to range selection when
the MAN / AUTO key is pressed, the rotary switch is activated or the
instrument is switched off and back on again.
Overview: Auto-Ranging and Manual Range Selection
* Via manual measuring range selection only
The multimeter is held in the selected measuring range. If the
range limit is exceeded, 0L appears at the display. You should then
switch to the next higher measuring range with the help of
the
key.
Measurement type MΩ
ISO
: If the measured value is less than 10% of
the measuring range after manual measuring range selection, u r
(under range) appears at the display. You should then select the
next smaller measuring range with the help of the
key.
Function Display
MAN / AUTO
Manual mode active:
utilized measuring range is fixed
MAN
or
Range switching sequence for:
V: 300 mV*
↔
3 V
↔
30 V
↔
300 V
↔
600 V
Hz: 300 Hz
↔
3 kHz
↔
30 kHz
↔
300 kHz (Hz(U))
Ω:
300
Ω ↔
3 k
Ω ↔
30 k
Ω ↔
300 k
Ω ↔
3 M
Ω ↔
30 M
Ω
A: 300 μA
↔
3 mA
↔
30 mA
↔
300 mA
↔
1 A
A : See chapter 5.7.3 und chapter 5.7.4
F :30 nF ↔ 300 nF ↔ 3 μF ↔ 30 μF ↔ 300 μF
MΩ
ISO
:
300 k
Ω ↔
3 M
Ω ↔
30 M
Ω ↔
300 M
Ω
F CAP: 300 nF ↔ 3 μF ↔ 10 μF
MAN
MAN / AUTO
Return to automatic measuring range selection —