Technical Data
Power Supply
Battery 2 ea. 1.5 V mignon cell (2 ea. size AA),
alkaline manganese per IEC LR6
Service life With alkaline manganese batteries: approx.
200 hours (without M
@UISO
measurement)
Battery test Battery capacity display with battery
symbol in 4 segments: .
Querying of momentary battery voltage via
menu function.
Power OFF function
The multimeter is switched off automatically:
– If battery voltage drops to below approx.
1.8 V
– If none of the keys or the rotary switch are
activated for an adjustable duration of 10
to 59 minutes, and the multimeter is not in
the continuous operation mode
Power pack socket
If the power pack has been plugged into
the instrument, the installed batteries are
disconnected automatically.
Rechargeable batteries can only be
recharged externally.
1)
Times 0.7 for interface operation
Electrical Safety
Safety class II per EN 61010-1:2001/VDE 0411-1:2002
Measuring category CAT II CAT III
Nominal voltage 1000 V 600 V
Fouling factor 2
Test voltage 5.2 kV~ per EN 61010-1:2001/VDE 0411-
1:2002
Fuses
Fuse link FF 10 A/1000 V AC/DC;
10 mm x 38 mm; switching capacity: 30 kA
at 1000 V AC/DC;
protects the current measurement input in
the 300 A through 10 A ranges
Electromagnetic Compatibility (EMC)
Interference emission
EN 61326-1:2006, class B
Interference immunity
EN 61326-1:2006
EN 61326-2-1:2006
Measuring
Function
Nominal
Voltage U
N
Resistance
of the
DUT
Service life
in Hours
Number of Possible
Measurements with
Nominal Current per
VDE 0413
V 200
1)
V 150
1)
M
100 V 1 M 50
100 V 100 k 3000
500 V 500 k 600
1000 V 2 M 200