, A/Hz
5.4 Continuity Test
Ð Disconnect supply power from the electrical circuit of the de-
vice to be measured, and discharge all high-voltage capaci-
tors.
Ð Make sure that the device under test is voltage-free.
Interference voltages distort measurement results!
Ð Set the rotary switch to “ ”.
Ð A loudspeaker symbol appears at the display.
Ð Connect the conductor path under test as shown.
Depending upon the selected limit value, the multimeter
generates a continuous acoustic signal in the case of continuity or
short-circuiting, i.e. at a value of less than the selected limit value.
“0L” appears at the display in the case of an open connection.
The limit value can be adjusted in the “SET ” menu (see also sec-
tion 6.4):
1nFo ... SET rAtE ... bEEP
1, 10, 20, 30, 40, 90
(10 = default setting)
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Temp
MISO
0V !
!
0.443
V
DC
R < 1, 10, 20, 30, 40, 90
FUNC
ENTER
R: 300
Measuring Range:
000.8
(0.1 to 310 )