User Manual

GMC-I Messtechnik GmbH 61
Technical Data
Electrical Safety
Safety class II per EN 61010-1:2001/VDE 0411-1:2002
Measuring category CAT II CAT III
Nominal voltage 1000 V 600 V
Fouling factor 2
Test v o l t a g e
5.2 kV~ per EN 61010-1:2001/VDE 0411-
1:2002
Fuses
Fuse link FF 10 A/1000 V AC/DC;
10 mm x 38 mm; switching capacity:
30 kA at 1000 V AC/DC;
protects the current measurement input in
the 300 A through 10 A ranges
Electromagnetic Compatibility (EMC)
Interference emission
EN 61326-1:2006, class B
Interference immunity
EN 61326-1:2006
EN 61326-2-1:2006
Data Interface
Type
Optical via infrared light through the housing
Data transmission Serial, bidirectional (not IrDa compatible)
Protocol Device specific
Baud Rate 38,400 baud
Functions Select/query measuring functions
and parameters
Query momentary measurement data
The USBX-TRA plug-in interface adapter (see accessories) is
used for adaptation to the PC’s USB port.
Internal Measured Value Storage
Memory capacity 4 MBit / 540 kB for approx. 15,000
measured values with indication of date
and time
Mechanical Design
Housing Impact resistant plastic (ABS)
Dimensions 200 x 87 x 45 mm
(without protective rubber cover)
Weight Approx. 0.35 kg with batteries
Protection Housing: IP 54
(pressure equalization by means of the
housing)
Table Excerpt Regarding Significance of the IP Code
IP XY
(1
st
digit X)
Protection against
foreign object entry
IP XY
(2
nd
digit Y)
Protection against the
penetration of water
4 1.0 mm dia. 4 splashing water
5 dust protected 5 water jets