Datasheet

GMC-I Messtechnik GmbH 5
METRAHIT ISO and METRAHIT COIL
TRMS Multimeter with Insulation Measurement
and Interturn Short-circuit Measurement (C
OIL only)
Response Time (after manual range selection)
Display
LCD panel (65 mm x 36 mm) with analog and digital display
including unit of measure, type of current and various special
functions
Background Illumination
Background illumination is switched off approximately 1 minute
after it has been activated.
Analog
Display LCD scale with pointer
Scaling Linear
:
5 ... 0 ... ±30 with 35 scale divisions
for , 0 ... 30 with 30 scale divisions in all
other ranges
Polarity display with automatic switching
Overflow display with the symbol
Measuring rate 40 measurements per second and display
refresh
Digital
Display / char. height 7-segment characters / 15 mm
Number of places 3¾ digits 3100 steps, the changeover
function to
digits
in measuring function
VDC and
Ω
depends on parameter selection
Overflow display “OL” is displayed for 30000 digits, or
3100 digits, respectively
Polarity display “–” (minus sign) is displayed
if plus pole is connected to “
Measuring rate 10 and 40 measurements per second with
the Min-Max function except for the
capacitance, frequency and duty cycle
measuring functions
Refresh rate 2 times per second, every 500 ms
Electrical Safety
Safety class
II per DIN EN 61010-1:2011/VDE 0411-
1:2011
Measuring category CAT II CAT III
Nominal voltage 1000 V 600 V
Pollution degree 2
Test voltage 5.2 kV~ per DIN EN 61010-1:2011/
VDE 0411-1:2011
Fuses
Fuse link FF 10 A / 1000 V AC/DC;
10 x 38 mm;
Switching capacity:
30 kA at 1000 V AC/DC,
protects the current measurement input in
the 300 μA through 10 A ranges
Power Supply
Battery 2 ea. 1.5 V mignon cell (2 ea. size AA),
alkaline manganese per IEC LR6
Service life With alkaline manganese batteries:
approx. 200 hours (without
MΩ
ISO
measurement)
Battery test Battery capacity display with battery sym-
bol in 4 segments: „ .
Querying of momentary battery voltage via
menu function.
Power OFF function
The multimeter is switched off automatically:
– If battery voltage drops to below
approx. 1.8 V
– If none of the keys or the rotary switch
are activated for an adjustable duration
(10 to 59 min.) and the multimeter is not
in the continuous operation mode
Power pack socket If the power pack has been plugged into
the instrument, the installed batteries are
disconnected automatically.
Rechargeable batteries can only be
recharged externally.
1
Times 0.7 for interface operation
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2013, class B
Interference immunity EN 61326-1:2013
EN 61326-2-1:2013
Ambient Conditions
Accuracy range 0 °C ... +40 °C
Operating temp. range 10 °C ... +50 °C
Storage temp. range 25 °C ... +70 °C (without batteries)
Relative humidity 40 to 75%, no condensation allowed
Elevation to 2000 m
Deployment
Indoors, except within specified ambient
conditions
Measured Quantity /
Measuring Range
Response Time,
Digital Display
Jump Function
of the Measured Quantity
V, V
A, A
1.5 s
From 0 to 80%
of upper range limit value
300 Ω ... 3 MΩ 2s
From to 50%
of upper range limit value
30 MΩ,
MΩ
@U
ISO
Max. 5 s
Continuity < 50 ms
°C (Pt 100) Max. 3 s
1.5 s
30 nF ... 300 μFMax. 5s
From 0 to 50%
of upper range limit value
>10 Hz 1.5 s
Measuring
Function
Nominal
Voltage U
N
Resistance
of the
DUT
Service Life
in Hours
Number of Possible
Measurements with
Nominal Current per
VDE 0413
V 200
1
V 150
1
MΩ
@U
ISO
100 V 1 MΩ 50
100 V 100 kΩ 3000
500 V 500 kΩ 600
1000 V 2 MΩ 200