User Manual
GMC-I Messtechnik GmbH 51
Device and Measuring Parameters
6.4 Entering Parameters – SETUP Menu
rAtE – set the sampling rate
The sampling rate specifies the time interval after which the
respective measured value is transmitted to the interface, or to
measured value memory.
Any one of the following sampling rates can be selected:
[mm:ss.t]: 00:00,1, 00:00,2, 00:00.5, 00:01,0, 00:02,0, 00:05.0
[h:mm:ss.t] (h=hours, m=minutes, s=seconds
, z=tenths of a sec.
):
0:00:10, 0:00:20, 0:00:30, 0:00:40, 0:00:50, 0:01:00, 0:02:00,
0:05:00, 0:10:00, 0:20:00, 0:30:00, 0:40:00, 0:50:00, 1:00:00,
2:00:00, 3:00:00, 4:00:00, 5:00:00, 6:00:00, 7:00:00, 8:00:00,
9:00:00
Setting the Sampling Rate
1nFo
...
SET rAtE
00:00.1 ... 00:00.5 ... 9:00:00
(00:00.5 = 0.5 s = default value)
rESoL – high resolution for V DC and (customer-specific function)
Switching back and forth between 3¾ and 4¾ places is possible
for direct voltage and resistance measurement.
1nFo
... SET rAtE ... resol
3000 / / 30000
(3000 = default setting)
0.diSP – show/hide leading zeros
This parameter determines whether or not leading zeros will
appear in the measured value display.
1nFo
...
SET rAtE ... 0.diSP
0000.0 : with leading zeros (default value)
0.0 : leading zeros suppressed
CLIP – set current clip factor
See chapter 5.8.3, chapter 5.8.4 and chapter 5.8.5.