User Manual

GMC-I Messtechnik GmbH 47
Measurements: V/Hz, , , Temp (°C/°F), , M
@UISO
, A/Hz
5.8.5 Alternating Current Measurement with A AC and Hz Clip-On
Current Transformer
Current/Current Transformer Output
When a clip-on current transformer is connected to the
multimeter (mA/A input), all current displays appear with the
correct value in accordance with the selected transformation
ratio. The only prerequisite is that the current transformer is
equipped with at least one of the below listed transformation
ratios, and that the ratio has been previously selected in the
following menu (Cl ip 0FF).
Current Clip Setup Menu
1nFo
... SET rAtE ... CLIP
1 / 10/100/1000 / 0ff

Transf. R atios
CL IP
DMM Measuring Ranges Clip Types
30 mA 300 mA 3 A
1:1
1mA / 1 mA
30.00 mA 300.0 mA 3.000 A
1:10
1 mA / 10 mA
300 mA 3.000 A 30.00 A
1:100
1 mA / 100 mA
3.000 A 30.00 A 300.0 A
1:1000
1 mA / 1 A
30.00 A 300.0 A 3000.0 A
WZ12A, WZ12D,
WZ11A, Z3511,
Z3512, Z3514
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Temp
MISO
~
R
x
Current
R
i
R
i
50
A~
Hz
A~
Hz
003
.
50
A
AC
TRMS
0050
.
1
Hz
AC
FUNC
ENTER
1:1000
1:1000
MEASURE
SETUP
... SET ... CL IP
FUNC
ENTER
FUNC
ENTER
 0ff
Measuring Ranges:
Hz: 1 Hz . . . 31 kHz
mA / A: see table
3 ranges: 300 Hz, 3 kHz and 30 kHz