Measurements: V/Hz, , , Temp (°C/°F), , M
@UISO
, A/Hz
5.6 Capacitance Measurement
➭
Disconnect supply power from the electrical circuit of the device
to be measured, and discharge all high-voltage capacitors.
➭ Make sure that the device under test is voltage-free.
Capacitors must always be discharged before measurement is
performed.
Interference voltages distort measurement results!
Refer to section 5.1.1 regarding testing for the absence of
voltage with the help of the direct voltage measurement.
➭ Set the rotary switch to “ ”.
➭ Connect the (discharged!) device under test to the sockets
with the measurement cables as shown.
Note
The “–” pole of polarized capacitors must be connected to the
“
” jack.
Resistors and semiconductor paths connected in parallel to
the capacitor distort measurement results!
COM mA A
V
Temp
MISO
COM mA A
V
Temp
MISO
0V !
!
10
.
38
F
+
–
+
–
Measuring range: 10 pF to 300 F
5 ranges: 30 nF, 300 nF,
3 F, 30 F and 300 F