Datasheet

GMC-I Messtechnik GmbH 3
METRAHITISO
TRMS Multimeter with Insulation Measurement
Technical Data
1
15 ... 45 ... 65 Hz ... 10 (5) kHz sine. See page 6 regarding influence
2
At 0 ... + 40 C
3
Display of up to max. 5.1 V, “OL” in excess of 5.1 V.
4
Applies to measurements at film capacitors and battery operated
5
Lowest measurable frequency for sinusoidal measuring signals symmetrical to the
zero point
6
Overload capacity of the voltage measurement input:
power limiting: frequency x voltage max. 3 x 10
6
V x Hz at > 100 V
7
Overload capacity of the current measurement input:
See current measuring ranges for maximum current values.
8
Input sensitivity, sinusoidal signal, 10% to 100% of voltage or current measuring
range; limitation: up to 30% of the range at up to 100 kHz in the mV measuring
range., 30% of the range in the 3 A measuring range
The voltage measuring ranges with max. 30 kHz apply in the A measuring range.
9
Plus sensor deviation
10
With ZERO function active
11
With short circuited terminal tips
Exception: residual value of 1 to 10 digits,
in the mV/A range
1 to 35 d at zero point due to the TRMS converter
12
10 minute cool-down period
Key: d = digit(s), MR = measuring range, rdg. = reading
Meas.
Func-
tion
(input)
Measuring Range
Resolution
at Upper Range Limit
Input Impedance
Intrinsic Uncertainty
under Reference Conditions
Overload
Capacity
2)
(... % rdg. + ... d)
30000 3000 3000 3000
30000 3000
~ /
~
1) 11)
1) 11)
Value Time
V
300.0 mV 10 V 100 V9M 9M // < 50 pF 0.15 + 15
10)
0.2 + 3
10)
1 + 3 (> 100 D) 1.5 + 5 (> 100 D)
1000 V
DC
AC
RMS
Sine
6)
Cont.
3.000 V 100 V1mV 9M 9M // < 50 pF 0.15 + 15 0.15 + 2
1 + 3 (> 30 D) 1.5 + 5 (> 100 D)
30.00 V 1 mV 10 mV 9 M 9M // < 50 pF 0.15 + 15 0.15 + 2
300.0 V 10 mV 100 mV 9 M 9M // < 50 pF 0.15 + 15 0.15 + 2
1000 V 100 mV 1 V 9 M 9M // < 50 pF 0.15 + 15 0.2 + 2
Voltage drop at approx. range limit
~
1) 11)
1) 11)
A
300.0 A 100 nA 18 mV 18 mV 0.5 + 5 1.5 + 5 (> 100 D) 1.5 + 5 (> 100 D)
0.3 A Cont.
3.000 mA 1 A 160 mV 160 mV 0.2 + 3
1.5 + 5 (> 30 D) 1.5 + 5 (> 100 D)
30.00 mA 10 A 32 mV 32 mV 0.5 + 3
300.0 mA 100 A 200 mV 200 mV 0.2 + 3
3.000 A 1 mA 120 mV 120 mV 1 + 5
10 A 5 min
12)
10.00 A 10 mA 400 mV 400 mV 1 + 5
Factor 1:1/10/100/1000
Input Input impedance
~
1) 11)
1) 11)
A
@ A
0.03/0.3/3/30 A 30 mA
Current measurement input
(jack A~)
1.5 + 5 (> 100 D) 0.3 A Cont.
0.3/3/30/300 A 300 mA
3/30/300/3k A
3 A Plus clip-on current transformer error 3 A 5 min
A
@ V
0.3/3/30/300 A 300 mV
Voltage measurement input approx. 9 M
(V socket)
0.5 + 3
1.5 + 3 (> 300 D) 1.5 + 5 (> 300 D) Meas. input
6)
:
3/30/300/3k A 3 V 1.5 + 3 (> 30 D) 1.5 + 5 (> 100 D)
1000 V
RMS
max. 10 s
30/300/3k/30k A
30 V Plus clip-on current sensor error
Open-circuit
voltage
Meas. current at
range limit
(... % rdg. + ... d)
30000 3000
300.0 10 m 100 m < 1.4 V Approx. 300 A
0.5 + 15
with ZERO active
0.5 + 3
with ZERO active
1000 V
DC
AC
RMS
Sine
max. 10 s
3.000 k 100 m 1 < 1.4 V Approx. 200 A 0.5 + 15 0.5 + 2
30.00 k 1 10 < 1.4 V Approx. 30 A 0.5 + 15 0.5 + 2
300.0 k 10 100 < 1.4 V Approx. 3 A 0.5 + 15 0.5 + 2
3.000 M 100 1k < 1.4 V Approx. 0.3 A 0.5 + 15 0.5 + 2
30.00 M 1k 10 k < 1.4 V Approx. 33 nA 2.0 + 20 2.0 + 5
300.0 100 m ca. 10 V
Approx. 1 mA const.
3 + 5
5.1 V
3)
1mV ca. 10 V 2 + 5
Discharge resist. U
0 max
(... % rdg. + ... d)
F
30.00 nF 10 pF 10 M 0.7 V 1 + 6
4)
with ZERO function activ
e
1000 V
DC
AC
RMS
Sine
max. 10 s
300.0 nF 100 pF 1 M 0.7 V 1 + 6
4)
3.000 F1nF100k 0.7 V 1 + 6
4)
30.00 F10nF12k 0.7 V 1 + 6
4)
300.0 F 100 nF 3 k 0.7 V 5 + 6
4)
f
min
5)
(... % rdg. + ... d)
Hz (V)/
Hz (A)
Hz (A )
300.0 Hz 0.1 Hz
1Hz
0.1 + 2
8)
Hz (V)
6)
:
Hz(A )
6)
:
1000 V
Hz (A):
7
)
max. 10 s
3.000 kHz 1 Hz
30.00 kHz 10 Hz 10 Hz
Hz (V)
300.0 kHz 100 Hz 100 Hz
(... % rdg. + ... d)
9)
C
Pt 100
200.0 ...
+850.0 C
0.1 C
0.5 %+ 15
1000 V
DC/AC
RMS
Sine
max. 10 sPt 1000
150.0 ...
+850.0 C
0.5 %+ 15
K
(NiCr-Ni)
250.0 ...
+1372.0 C
1% + 5 K