User Manual
GMC-I Gossen-Metrawatt GmbH 59
Technical Data
Influencing Quantities and Influence Error 
1)
With zero balancing
2)
Power limiting: frequency x voltage max. 6 x 10
6
 V x Hz for U > 100 V
3)
The accuracy specification for frequency response is valid within a dis-
play value range of 10% to 100% of the measuring range for both mea-
suring modes with the TRMS converter in the AC and (AC+DC) ranges.
4)
METRAHIT EXTRA:  Frequency response up to 50 kHz, 
METRAHIT E
TECH:  Frequency response up to 20 kHz, 
METRAHIT E
SPECIAL:  Frequency response up to 20 kHz, 
METRAHIT E
BASE:  Frequency response up to 1 kHz
Influencing 
Quantity
Sphere of 
Influence
Measured Quantity /
Measuring Range 
1)
Influence Error 
(...% rdg. + ... d) / 10 K
Temperature
0 °C ... +21 °C
and
+25 °C ... +40 °C
V 0.2 + 10
V 0.4 + 10
600 Ω ... 6 MΩ 0.5 + 10
> 6 MΩ 1 + 10
mA/A    0.5 + 10
mA/A  0.8 + 10
60 nF ... 600 μF1 + 5
Hz, dB 0.2 + 10
°C/°F (Pt100/Pt1000) 0.5 + 10
°C/°F thermocouple K 0.2 + 10
Influencing 
Quantity
Measuring Quantity
Influence Error
(...% rdg. + ... d)
DATA V, A, Ω, Hz, dB, °C ±10 d
MIN / MAX V, A, Ω, Hz, dB, °C ±30 d
Influencing 
Quantity
Meas. Quantity/
Meauring Range
Sphere of Influence
Intrinsic Error 
3) 
±( ... % rdg. + ... d)
METRAHIT E
XTRA
METRAHIT E
TECH
METRAHIT E
SPECIAL
METRAHIT E
BASE
Frequency
V
AC
600.00 mV
> 15 Hz  ... 45 Hz 3 + 30 3 + 30
> 65 Hz ... 1 kHz 2 + 30 3 + 30
> 1 kHz ... 20 kHz  3 + 30 —
6.0000 V
... 
600.00 V
 2)
> 15 Hz  ... 45 Hz 2 + 9 3 + 9
> 65 Hz ... 1 kHz 1 + 9 3 + 9
> 1 kHz ... 20 kHz 
4)
3 + 9 —
> 20 kHz ... 50 kHz 
4)
—
1000.0 V
2) 
> 15 Hz  ... 45 Hz 2 + 9 3 + 9
> 65 Hz ... 1 kHz 2 + 9 3 + 9
  > 1 kHz ... 10 kHz  3 + 30  —
A
AC
600.00 μA
...
10.0000 A
> 15 Hz ... 45 Hz
3 + 10 —
> 65 Hz  ...  10 kHz
A
AC 
E
BASE
600 mV / 
6 V / 10 V
> 65  Hz  ...  1  kHz — 3 + 30