User Manual
GMC-I Gossen-Metrawatt GmbH 51
Device and Measuring Parameters
0.diSP – show/hide leading zeros
This parameter determines whether or not leading zeros will 
appear in the measured value display.
1nFo
... 
SET  rAtE  ...  0.diSP   
0000.0 : with leading zeros (default setting)
0.0 : leading zeros suppressed 
 
A.diSP – analog display: select display mode
One of two different display modes can be selected for the 
analog display:
• bArG: bar graph
• Point: pointer
 1nFo 
 ...  SET  rAtE  ...  A.diSP
 bArG / Point 
 
(bArG = default value)
APoFF – specified time for automatic shutdown and continuous ON
The instrument is switched off automatically if the measured value 
remains unchanged for a long period of time, and if none of the 
keys or the rotary switch have been activated before the specified 
time “APoFF ” (entered in minutes) has elapsed.
If the on setting is selected, the multimeter is set to continuous 
operation and 
ON appears in the display to the right of the battery 
symbol. In this case, the multimeter can only be switched off 
manually. The „on “ setting can only be cancelled by changing the 
respective parameter, or by switching the instrument off manually. 
In this case, the parameter is reset to 10 minutes.
 1nFo 
 ...  SET  rAtE  ...  APoFF
 10 ... 59 min on 
   
(10 minutes = default setting)
dbrEF – measuring alternating voltage level
 1nFo 
 ...  SET   rAtE  ...  dbrEF
 0 ..001 ... 99 V 
   
(0,775 V = default setting)