User Manual
46 GMC-I Gossen-Metrawatt GmbH
V/Hz, Ω, Temperature,   and A/Hz Measurements
5.7.5
Direct, Pulsating and Current Measurement
 with Current Clamps 
Transformer A DC, A (DC+AC), A AC and Hz (METRAHIT E
XTRA, 
METRAHIT E
TECH and METRAHIT ESPECIAL only)
Transformer Output, Current/Current
When a current clamps transformer is connected to the multime-
ter (   mA/A input), all current displays appear with the correct 
value in accordance with the selected transformation factor. The 
only prerequisite is that the current transformer is equipped with 
at least one of the below listed transformation factors, and that 
the factor has been previously selected in the following menu 
(
Cl
ip
≠ 
0FF
)
, see also chapter 6.4.
Current Clamp Setup Menu
 1nFo 
 ...  SET   rAtE  ...  CLIP 
 1 / 10/100/1000 / 0ff 
 
(default setting: Cl ip = 0FF)
Trans. Factors
CL IP
Measuring Ranges DMM Clamp Types
60 mA AC 600 mA AC 6 A AC
1:1 
1mA/1mA
60.000 mA 600.00 mA 6.0000 A
WZ12A, WZ12D, 
WZ11A, Z3511, 
Z3512, Z3514
1:10 
1mA/10mA
600.00 mA 6.0000 A 60.000 A
1:100 
1mA/100mA
6.0000 A 60.000 A 600.00 A
1:1000 
1 mA/1 A
60.000 A 600.00 A 6000.0 A