User Manual
44 GMC-I Gossen-Metrawatt GmbH
V/Hz, Ω, Temperature,   and A/Hz Measurements
5.7.3 Direct and Pulsating Current Measurement 
with Current Clamp Sensors, A DC and A (DC+AC)
Transformer Output, Voltage/Current
When a current clamp sensor is connected to the multimeter 
(V input (
METRAHIT EBASE
:  V input)), all current displays appear with 
the correct value in accordance with the selected transformation fac-
tor. The only prerequisite is that the current sensor is equipped with at 
least one of the below listed transformation factors, and that the fac-
tor has been previously selected in the following menu (Cl ip ¼ 0FF), 
see also chapter 6.4.
Current Clamp Setup Menu
 1nFo 
 ...  SET   rAtE  ...  CLIP 
 1 / 10/100/1000 / 0ff 
 
Maximum allowable operating voltage is equal to the current trans-
former’s nominal voltage. When reading the measured value, addi-
tional error resulting from the current clamp sensor must also be 
taken into consideration.
(default setting: 
METRAHIT E
XTRA
, 
METRAHIT E
TECH
, 
METRAHIT E
SPECIAL
: 
OFF
, 
METRAHIT E
BASE
: 
1:1000
)
Trans. Factors
CL IP
Measuring Ranges DMM Clamp Types
600 mV 6 V
1:1 
1mV/1mA
600.00 mA 6.0000 A
1:1 
1mV/1mA
1:10 
1mV/10mA
6.0000 A 60.000 A
1:10 
1mV/10mA
1:100 
1mV/100mA
60.000 A 600.00 A
1:100 
1mV/100mA
1:1000 
1 mV/1 A
600.00 A 6000.0 A
1:1000 
1 mV/1 A
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM
V
Temp
Ω
A
~
R
x
Current
003
.
50
A
1:1000
A
DC
003
.
50
A
1:1000
A
DC
AC
TRMS
A
A
FUNC
ENTER
R
i
R
i
 = 9 M
Ω
COM
V
Temp
Ω
V
~
R
x
Hz
Current
R
i
R
i
 ~ 1 M
Ω
METRAHIT EBASE
mA
METRAHIT EBASE
METRAHIT EXTRA,
V
V
METRAHIT ETECH,
METRAHIT E
SPECIAL
METRAHIT EXTRA,
METRAHIT E
TECH,
METRAHIT E
SPECIAL